T1 - A Prospective Comparison of Standard Technique, Doppler Ultrasonography, and Pressure Waveform Analysis for Confirming Intraosseous Catheter Placement
AU - Greenstein, Yonatan Y.
AU - Ferrara, Scott
AU - Jomi, Jerry
AU - Soetanto, Vanessa
AU - Zuckerman, Omry
AU - Sena, Ariel
AU - Ostrovsky, Ilya
AU - Alerhand, Stephen
N1 - Publisher Copyright:
© The Author(s) 2023.
PY - 2024/3
Y1 - 2024/3
N2 - Purpose: Intraosseous (IO) catheters allow healthcare workers to rapidly administer fluids and medications to critically ill patients when intravenous access is inadequate or unable to be obtained. An improperly placed IO catheter can lead to delays in care, as well as serious complications such as limb necrosis. Methods: In this single-center, prospective, observational study, we compared 2 established methods of confirming proper IO catheter placement to a novel pressure waveform analysis technique in which the IO catheter is attached to a standard pressure transducer. Attaching a pressure transducer to a properly placed IO catheter produces a pulsatile waveform. Misplacement of the IO catheter produces a flatline waveform. Results: Of 42 IO catheters, 8 (19%) were incorrectly placed per the waveform analysis technique. Compared to the pressure waveform analysis technique, the standard method and the power Doppler method incorrectly classified 4/8 (50%) and 5/8 (62.5%) of the misplaced catheters, respectively. The standard method had a higher positive predictive value for detecting incorrectly placed IO catheters than the power Doppler method (100% vs 63%, respectively). Blinded reviewers demonstrated better agreement using the pressure waveform analysis technique than using power Doppler (k = 0.77 vs k = 0.58, respectively). Conclusion: The standard and power Doppler ultrasonography techniques identify incorrectly placed IO catheters sub-optimally. The pressure waveform analysis technique is more accurate than the standard of care and has superior interrater agreement compared to the ultrasound method of confirmation. With more than 500 000 IO catheters placed in the United States each year, this novel technique may improve overall IO safety. Trial Registration Number: NCT03908879.
AB - Purpose: Intraosseous (IO) catheters allow healthcare workers to rapidly administer fluids and medications to critically ill patients when intravenous access is inadequate or unable to be obtained. An improperly placed IO catheter can lead to delays in care, as well as serious complications such as limb necrosis. Methods: In this single-center, prospective, observational study, we compared 2 established methods of confirming proper IO catheter placement to a novel pressure waveform analysis technique in which the IO catheter is attached to a standard pressure transducer. Attaching a pressure transducer to a properly placed IO catheter produces a pulsatile waveform. Misplacement of the IO catheter produces a flatline waveform. Results: Of 42 IO catheters, 8 (19%) were incorrectly placed per the waveform analysis technique. Compared to the pressure waveform analysis technique, the standard method and the power Doppler method incorrectly classified 4/8 (50%) and 5/8 (62.5%) of the misplaced catheters, respectively. The standard method had a higher positive predictive value for detecting incorrectly placed IO catheters than the power Doppler method (100% vs 63%, respectively). Blinded reviewers demonstrated better agreement using the pressure waveform analysis technique than using power Doppler (k = 0.77 vs k = 0.58, respectively). Conclusion: The standard and power Doppler ultrasonography techniques identify incorrectly placed IO catheters sub-optimally. The pressure waveform analysis technique is more accurate than the standard of care and has superior interrater agreement compared to the ultrasound method of confirmation. With more than 500 000 IO catheters placed in the United States each year, this novel technique may improve overall IO safety. Trial Registration Number: NCT03908879.
KW - critical care
KW - intraosseous access
KW - resuscitation
KW - vascular access
UR - http://www.scopus.com/inward/record.url?scp=85170293064&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85170293064&partnerID=8YFLogxK
U2 - 10.1177/08850666231199086
DO - 10.1177/08850666231199086
M3 - Article
C2 - 37647305
AN - SCOPUS:85170293064
SN - 0885-0666
VL - 39
SP - 222
EP - 229
JO - Journal of Intensive Care Medicine
JF - Journal of Intensive Care Medicine
IS - 3
ER -