Incomplete deployment of the Vena Tech LP filter-case series and concerns

Sohail Contractor, Azadeh Esmaeili, Diego Reina, Edwin Deitch

Research output: Contribution to journalArticle

4 Scopus citations

Abstract

The Vena Tech LP vena cava filter (B Braun, Evanston, Illinois) has been FDA approved since 2001 and is a permanent vena cava filtration device. It replaced the previous Vena tech LGM filter also manufactured by B Braun. The LGM filter had 2 case series reporting a high incidence of incomplete deployment of the filter, especially when placed from a jugular approach. Design changes were made to this device and the LP filter introduced. The LP filter has also been reported to have incompletely deployed both in peer reviewed literature as well as the FDA MAUDE website. We present here 3 cases of incomplete deployment of the Vena Tech LP filter and review the cases previously described as well as attempt to present possible etiologies for incomplete deployment.

Original languageEnglish (US)
Pages (from-to)345-351
Number of pages7
JournalVascular and Endovascular Surgery
Volume45
Issue number4
DOIs
StatePublished - May 1 2011

    Fingerprint

All Science Journal Classification (ASJC) codes

  • Surgery
  • Cardiology and Cardiovascular Medicine

Cite this